Equation:

Pressure Change = ((Fx8xLxV)/(pi x r4)) x 2.42 x 10-9

  • F = Flow Rate in mL/minute
  • L = Length of Tubing
  • V= Vicsocity of Mobile Phase in Centipose
  • r = Radius of Tubing (Internal)