UV spectrophotometer & UV detector calibration | HOX solutions.
MicroSolv's Chemical Solutions™ branded UV-vis calibration kit contains everything you need to perform all the tests necessary for a proper performance qualification (PQ) and calibration of your UV-vis spectrophotometer or an HPLC UV detector. Use all or only part of the kit to adapt to your current SOP or create a new one using the included background information. A series of NIST traceable solutions and software designed to allow the laboratory user to qualify a UV-vis spectrophotometer easily and since it is fully DIY, that makes it easy to properly qualify your instrument when you need while saving a lot of money.
Test any UV or visible wavelength detector / spectrophotometer for the following performance characteristics:
Absorbance accuracy, absorbance linearity, photometric precision, wavelength accuracy, wavelength precision, stray light (based on NIST SRM 2032).
Click HERE for more information on the UV-vis calibration kit
Click HERE for specifications.
Kits sold will have at least six months until the expiration date. They are designed to be used promptly and often. Expiration date is set at two years after the date of manufacture.